Philips BiPAP - Flow Rate Data and Contact Information; Maximum Flow Rate Data; Contacting Respironics Support

Philips BiPAP
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 User Manual
Maximum Flow Rate (typical)
Auto Bi-level/Bi-level:
Test pressures (cm H
2
O)
4.0 9.0 14.5 20.0 25.0
22 mm
tubing
Measured pressure at the patient
connection port (cm H
2
O)
3.6 8.5 13.5 19.0 24.1
Average ow at the patient
connection port (l/min)
84.1 145.2 153.9 128.7 138.9
15 mm
tubing
Measured pressure at the patient
connection port (cm H
2
O)
3.8 8.0 13.5 19.0 24.0
Average ow at the patient
connection port (l/min)
85.1 122.3 120.6 119.2 138.5
CPAP:
Test pressures (cm H
2
O)
4.0 8.0 12.0 16.0 20.0
22 mm
tubing
Measured pressure at the patient
connection port (cm H
2
O)
3.6 7.5 11.0 15.0 19.0
Average ow at the patient
connection port (l/min)
84.1 135.2 154.5 146.9 128.7
15 mm
tubing
Measured pressure at the patient
connection port (cm H
2
O)
3.8 7.0 11.0 15.0 19.0
Average ow at the patient
connection port (l/min)
85.1 120.7 121.6 119.3 119.2
Disposal
Separate collection for electrical and electronic equipment per EC Directive 2002/96/EC. Dispose of this device in
accordance with local regulations.
How to Contact Respironics
To have your device serviced, contact your home care provider. If you need to contact Respironics directly, call the
Respironics Customer Service department at 1-800-345-6443 or 1-724-387-4000. You can also use the following