Set up device
Provider setup menu (cont.)
Alarm settings
Apnea 0, 10, 20, 30 sec
Patient disconnect 0, 15, 60 sec
Low min vent 0 (off) to 99 lpm in 1.0 increments
Low Vte 0 (off) 1 (on)
System settings
Backlight On, Off
Language English, Italian, Spanish, German, French, icons
hPa/cm H
2
O hPa, cm H
2
O
Setup parameter display
1
Leak, RR, Vm, Vte
Humidifier 0 to 5
Reset blower hours Yes, No
Provider mode
2
On, Off
Provider setup menu
Therapy settings
Mode CPAP, S, S/T, T, PC
CPAP 4-20
AVAPS
(not available on BiPAP S/T)
Yes, No
Vte 200-1500
IPAP max 4-25
IPAP min 4-25
IPAP 4-25
EPAP 4 to IPAP
BPM 0-30
Ti 0.5-3.0
Comfort settings
Flex control Yes, No
Flex setting Off, 1, 2, 3
Rise time control Yes, No
Rise time 1-3
Ramp 0-45 min
Ramp start 4 to EPAP
1. Setup parameter display
You can select which measured parameters will display
on the Monitor Pressure screen. Choose from Leak, RR
(Respiratory Rate), Min vent, or Vte. See the Measured
Parameter section of the User Manual for more information.
2. Provider mode
You can choose onor offto enable or disable provider
mode. Selecting onwill put the device in provider mode
(and the device will not automatically return to patient mode
when the screen times out or you select the Back option).
Selecting nokeeps the device in patient mode.
To access provider settings:
1. From the main menu, highlight Setup.
2. Press and hold ramp button and control wheel
simultaneously for 4-5 seconds until beep.
3. Provider Setup screen then appears.
To program device:
1. Turn control wheel to specific setting and press.
2. Turn control wheel to desired option and press control
wheel to select.
3. Turn control wheel to next setting and repeat.
4. To exit, turn control wheel to Back and press.
BiPAP AVAPS
BiPAP S/T
Quick start guide
• Set up device
• Start therapy
• View information
• Comfort settings
• Undocking
• Humidification
• Humidifier settings
Please visit http://bipapavaps.respironics.com or http://bipapst.respironics.com
Philips Healthcare is part of
Royal Philips Electronics
How to reach us
www.philips.com/healthcare
healthcare@philips.com
fax: +31 40 27 64 887
Asia
+852 2821 5888
Europe, Middle East, Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585
(toll free, US only)
Philips Respironics
1010 Murry Ridge Lane
Murrysville, PA 15668
Customer Service
+1 724 387 4000
800 345 6443
(toll free, US only)
Respironics Asia Pacific
+65 6882 5282
Respironics Australia
+61 (2) 9666 4444
Respironics Deutschland
+49 8152 93 06 0
Respironics Europe,
Middle East, Africa
+33 1 47 52 30 00
Respironics France
+33 2 51 89 36 00
Respironics Italy
+39 039.203.1
Respironics Sweden
+46 8 120 45 900
Respironics Switzerland
+41 6 27 45 17 50
Respironics United Kingdom
+44 800 1300 845
www.philips.com/respironics

Other manuals for Philips BiPAP AVAPS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips BiPAP AVAPS and is the answer not in the manual?

Philips BiPAP AVAPS Specifications

General IconGeneral
ModelBiPAP AVAPS
CategoryHumidifier
Humidification TechnologyHeated Humidification
Adjustable Humidity LevelsYes
BrandPhilips
Water Chamber Capacity325 ml
Heating ElementIntegrated
Power SourceElectric
MaintenanceRegular cleaning

Related product manuals