EasyManua.ls Logo

Philips DreamStation CPAP - Device Settings and Features; Connectivity and Mask Fit Check

Philips DreamStation CPAP
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13User Manual
Icon Text Description
Mask Type This setting allows you to adjust the level of air pressure relief based on the
specic Philips Respironics mask. Each Philips Respironics mask may have a
System One” resistance control setting. Contact your home care provider if
you cannot nd this resistance setting for your mask.
Note: If a lock icon is displayed on this screen, it indicates that your
provider has locked this setting and you cannot change it.
Tube Type This setting allows you to select the correct size diameter tubing that you are
using with the device. You can choose either (22) for the Philips Respironics 22
mm tubing, or (15) for the Philips Respironics 15 mm tubing. When using Heated
Tubing, the device will automatically change this setting to the appropriate tubing
type (15H) and you will not be able to change it.
Note: Tubing is identied on the cuff with the tubing identier symbol: “15”,
“22” or “15H”.
Note: If a lock icon is displayed on this screen, it indicates that your
provider has locked this setting and you cannot change it.
Language This feature allows you to choose which language to display on the interface.
You can choose between: English, German, Spanish, French, Italian, Brazilian
Portuguese, Danish, Finnish, Norwegian, Swedish, Czech, or Polish. You can also
turn the language off (0), which means the device will only display icons on the
interface.
Check Mask Fit This feature allows you to check the t of your mask prior to starting therapy.
This is done by measuring the amount of leak.
Modem Allows you to turn modem off temporarily or turn it back on. When modem
is turned off, it will automatically turn on again after 3 days. Only displays when
modem is installed.
Bluetooth Allows you to turn Bluetooth off and on. Also, it allows you to clear the pairing
with a compatible Bluetooth device.
Time Allows you to adjust the time. The default setting is Greenwich Mean Time, but
you may adjust the time in 30 minute increments to match your local time zone.
Note: This time setting is not displayed as a clock function on the device. It is
only used to align your therapy data for your Provider’s data reports.

Related product manuals