16 User Manual
Device Alerts
Device alerts are pop-ups that show up on the UI screen. There are 5 types of alerts described here:
Status: These alerts are just the pop-up screen.
Notication: These alerts consist of the pop-up screen in addition to a blinking Power LED on top of the device.
Alert 1: These alerts consist of the pop-up screen, a blinking Power LED and an audible beep when displayed. This
alert will not occur during therapy.
Alert 2: These alerts consist of the pop-up screen, a blinking Power LED and an audible beep when displayed. This
alert can occur during therapy.
Safe State: These alerts consist of the pop-up screen, a blinking Power LED and a repeating audible beep.
Note: Status alerts automatically time out after 30 seconds and their pop up screens disappear. All other alerts
must be acknowledged to clear.
Alert Summary Table: The following table summarizes the alerts.
Alert Icon Type Description Possible Cause Action
Data Activity: Do
not remove SD
card.
Status SD card read/write
underway.
n/a No action needed.
Change Accepted
Status Conrms acceptance
of prescription change
or device upgrade.
n/a No action needed.
EZ-Start
Pressure
Incremented to
xx.x
Status Displays when
EZ-Start mode is
enabled and device
is increasing therapy
pressure setting for
the next session.
n/a No action needed.
Oximetry: Good
Connection
(icon only)
Status Displays on the
therapy screen when
the blower is on
and 3 seconds of
good connection is
detected. Appears
at the beginning of
therapy. This screen
will not display
again if the nger
probe is removed
and reapplied unless
therapy is stopped
and restarted.
n/a No action needed.
Pair?: 123456
Yes/No
Status Prompts to accept or
decline pairing to a
Bluetooth compatible
device. This device
can be identied by
the digits displayed.
n/a Rotate control dial to
accept pairing (Yes),
or decline (No), then
press control dial to
conrm selection.