EasyManua.ls Logo

Philips EarlyVue VS30 - Page 127

Philips EarlyVue VS30
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EarlyVue VS30 Instructions for Use 5-7
Changing SpO
2
Settings
Changing the SpO
2
Sensitivity Mode
The Sensitivity Mode setting determines how quickly the monitor reports changes in
SpO
2
values by controlling the averaging algorithm used to calculate the SpO
2
value.
Note
If your system administrator has specified an SpO
2
alarm delay time, consider the total
possible delay time when selecting the Sensitivity Mode setting. For more information,
see “SpO
2
Alarm Delay” on page 5-9.
To change the SpO
2
sensitivity mode
1 Touch the SpO2 numeric pane.
The SpO2 menu appears, displaying the current settings.
2 Touch the Sensitivity Mode menu and select one of the following:
- Slow — Use this setting when motion artifact is an issue. SpO
2
changes are
reported every 20 seconds.
- Normal — Use this setting for most monitoring situations. Changes are reported
every 10 seconds.
- Fast — Use this setting for special applications (for example, sleep studies) when
you need a fast response. Changes are reported every 5 seconds. Do not use the
Fast setting if motion artifact is an issue.
3 Touch Done.
Depending on the magnitude of difference between the alarm limit and the displayed
value, the alarm signal generation delay may be from 1 second to the value of the
response time (5, 10, or 20 seconds), plus any additional configured alarm delay time. If
an SpO2 Extd Update technical alarm occurs, the physiological alarm response time is
proportionately longer. If the monitor is configured for SpO2 High/Low Alarm Delay, the
alarm generation is delayed by the configured time.
If the total delay in response time exceeds 30 seconds, the SpO2 numeric pane
alternates between displaying the SpO
2
value and displaying -?-.
Changing the SpO
2
Alarm Limits
For information on changing the SpO
2
alarm limits, see “To change the alarm limits for
one measurement” on page 4-8.
The SpO
2
default alarm limits are:
Adult Pediatric Neonatal
SpO
2
high limit 100% 100% 95%
SpO
2
low limit 90% 90% 85%

Related product manuals