Changing Temperature Settings
7-4 EarlyVue VS30 Instructions for Use
Note
If the monitor is in Standby mode and you press the On/Standby button to turn the
monitor on, the temperature module takes several seconds to reset. Do not remove the
temperature probe from the well while the module resets.
Changing Temperature Settings
Use the Temperature menu to change:
Temperature source
Temperature unit of measure
Temperature mode
Temperature probe sites
Temperature alarm audio on or off
Temperature alarm limits
When changing settings, make sure you follow the policies of your healthcare facility.
To open the Temperature menu, touch the Temp numeric pane. The Temperature menu
opens, displaying the current settings.
Changing the Temperature Source
The values displayed in the Temp numeric pane can be:
Sent to the VS30 monitor from a temperature probe connected to the monitor.
Manually entered, if you use a thermometer not connected to the monitor.
If you select Manual Input as the source, the following occurs:
A new field, named Input Temperature Measurement, appears on the Temperature
menu, enabling you to manually enter the temperature value.
The alarm limits fields disappear from the Temperature menu.
The Alarm Audio On/Off button disappears from the Temperature menu.
In the Temp numeric pane:
- The No Alarm icon is displayed to indicate that no visual or auditory alarm
indicators will be triggered for temperature.
- The phrase Manual Input is displayed to indicate that the temperature value is
sourced from manual input.
- The measurement value appears in gray.