EarlyVue VS30 Instructions for Use 14-5
Cleaning and Disinfecting the Predictive Temperature Module
Cleaning and Disinfecting the Predictive Temperature Module
Caution
When cleaning the temperature module, follow these general precautions:
Do not use steam, heat, or gas sterilization on the probe or probe well.
Do not autoclave the probe or probe well.
Cleaning and Disinfecting the Probe and Cord
To clean the probe and cord
Dampen a soft cloth with mild soap and warm water and wipe the probe and cord.
To disinfect the probe and cord
Dampen a soft cloth with any one of the following solutions and wipe the probe and
cord:
- Isopropyl alcohol (70% solution in water)
- Sodium hypochlorite (chlorine bleach, 10% solution in water)
- A nonstaining disinfectant
Cleaning and Disinfecting the Probe Well
To clean the probe well
1 Disconnect the probe and remove it from the well.
2 Remove the well from the monitor.
3 Dampen a soft cloth with mild soap and warm water and wipe the inner and outer
surfaces.
4 Thoroughly dry all of the surfaces before replacing the well in the monitor.
To disinfect the probe well
1 Disconnect the probe and remove it from the well.
2 Remove the well from the monitor.
3 Dampen a soft cloth with any of the following solutions and wipe the inner and outer
surfaces:
- Isopropyl alcohol (70% solution in water)
- Sodium hypochlorite (chlorine bleach, 10% solution in water)
- A nonstaining disinfectant
4 Thoroughly dry all of the surfaces before replacing the probe well in the monitor.