Preventive Maintenance
14-8 EarlyVue VS30 Instructions for Use
The following hand cleansers are acceptable for use with the HS-1 scanner:
Purell w/65% Alcohol
3M Avagard D
EcoLab Quik-Care
Preventive Maintenance
Preventive maintenance and repair should be performed only by trained service
personnel.
Mild detergent Per manufacturer’s recommendation, as
needed
Expose 256 0.5 oz/gallon water
Super Sani-Cloth Not approved
Virkon Per manufacturer’s recommendation
Warm water Not approved
Wexcide 1 oz/gallon water
Wexcide-Ready-To-Use No dilution required
Cleaning Agents Dilution
Suggested Testing Frequency
NBP calibration Once every two years.
Temporal temperature calibration
1
Once a year.
CO
2
calibration First calibration at 1,200 hours of use
or after one year, whichever comes
sooner.
After the first calibration, once a year
or after 4,000 hours, whichever comes
sooner.
After any repairs or the replacement of
any parts.
Replace the entire CO
2
module after
20,000 hours of use.
Performance
Predictive temperature accuracy
NBP accuracy test
•S
P
O
2
Once every two years, or if you suspect
the measurement is incorrect.