Safety Standards
16-2 EarlyVue VS30 Instructions for Use
Safety Standards
Electrical Specifications
Note
The only supported battery for the EarlyVue VS30 is the Philips lithium-ion battery with
Part Number 989803199221, listed in “Miscellaneous Accessories” on page 15-11. The use
of any unsupported (or unrecognized) battery may result in longer charging times and
unpredictable performance.
Parameter Specification
IEC 60601-2-49: 2011, IEC 60601-1:2005+A1:2012, IEC 60601-1-2: 2014, IEC 60601-1-
6:2010+A1:2013 (including IEC 62366: 2007 + A1: 2014), IEC 62304:2006+A1:2015, IEC
60601-1-8: 2006/AMD1:2012, IEC 60601-1-9: 2013, IEC 80601-2-30:2009+A1:2013,
ISO 80601-2-55: 2011, ISO 80601-2-56: 2017, ISO 80601-2-61: 2011
Protection Class Class I, internally powered equipment, per
IEC 60601-1
Degree of Protection Type CF defibrillator-proof, per IEC 60601-1
Mode of Operation Continuous
Protection Against Hazards of
Ignition of Flammable Anesthetic
Mixtures
Equipment is not suitable for use in the presence
of a flammable anesthetic mixture with air or
oxygen or nitrous oxide, per IEC 60601-1
Parameter Specification
Internal Battery Lithium ion, 9-cell smart battery
Output voltage: 11.1 V
Typical charge capacity: 7800 mAh
Complies with Smart Battery Data Specification (SMBus
v1.1)
Battery Operating Time
(new, fully-charged
battery)
At least 6 hours of monitoring with continuous SpO
2
and
NBP measured every 15 minutes at 25° C (77° F)
Total battery capacity decreases with age.
Battery Charge Time (fully
discharged battery)
< 5 hours with continuous SpO
2
and NBP measured
every 15 minutes at 25° C (77° F)
< 4 hours in Off or Standby mode