EarlyVue VS30 Instructions for Use 16-3
Environmental Specifications
Environmental Specifications
Caution
The monitor might not meet performance specifications if stored or used outside the specified
temperature and humidity ranges.
Internal Power Supply 100 VAC–240 VAC line voltage, with minimum tolerance
of ±10%
Provides secondary voltages <20 VDC and supports at
least 30 W of load.
Maximum Output Power
Consumption
60 W
Frequency 50/60 Hz, with minimum tolerance of ±5%
Parameter Specification
Parameter Specification
Mechanical Shock Complies with the mechanical shock requirement in ISO
80601-2-61 standards for use within the healthcare facility.
Test conditions include:
Peak Acceleration: 150 m/s
2
(15 g)
Duration: 11 ms
Pulse shape: half sine
Number of shocks: 3 shocks per direction per axis (18 total)
Mechanical Vibration Complies with the mechanical vibration requirement in ISO
80601-2-61 standards for use within the healthcare facility.
Test conditions include:
• Frequency range: 10 Hz to 2000 Hz
• Resolution: 10 Hz
• Acceleration amplitude:
- 10 Hz to 100 Hz: 1.0 (m/s
2
)
2
/Hz
- 100 Hz to 200 Hz: –3.0 db/octave
- 200 Hz to 2000 Hz: 0.5 (m/s
2
)
2
/Hz
Duration: 10 min per each perpendicular axis (3 total)