EasyManua.ls Logo

Philips EarlyVue VS30 - Page 249

Philips EarlyVue VS30
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EarlyVue VS30 Instructions for Use 16-7
CO2 Specifications
To determine the accuracy of airway respiration rate (awRR), Microstream™ technology
simulates breaths for use in respiration rate measurement with a system that uses a tank
of N
2
(representing no CO
2
for inhalation) and a tank of CO
2
(of the percent CO
2
required
for the particular test). A control board, triggered by a computer, uses solenoids to switch
the module input between the two tanks of gas, creating a gas CO
2
square wave. This
system can create simulated breaths over the full required range of specified respiration
rates.
Parameter Specification
CO
2
units mmHg
etCO
2
and
imCO
2
Measurement Range
0 mmHg–150 mmHg
Data Sample Rate Waveform sampling, 20 samples per second
Flow Rate 50 ml/min, + 15 ml/min, –7.5 ml/min, flow measured by
volume.
1
CO
2
Reporting
Resolution
0.1 mmHg
etCO
2
and
imCO
2
Displayed
Resolution
1 mmHg
Initialization Time The time between powering up the monitor and presenting
the first reading is typically 40 seconds; maximum 3 minutes
CO
2
Response Time The total response time for adults/pediatrics is
approximately 2.9 seconds, for 10% to 90% changes in CO
2
concentration. Calculated as follows:
Delay time + Rise time + System response time
CO
2
Delay time: 2.7 seconds (typical)
CO
2
Rise time: 0.2 seconds (typical)
Monitor response time: 1.0 second
The maximum CO
2
response time (with a standard-length
FilterLine™) is 5.3 seconds (typical).
Adult/Pediatric Long FilterLines and Long Smart CapnoLines
have a maximum CO
2
response time of 6.9 seconds (typical).
For information on CO
2
response time for infant/neonatal
patients, see “CO2 Response Time for Infant/Neonatal
Accessories” on page 16-9.
Calibration Interval First calibration at 1,200 hours of use or after one year,
whichever comes sooner.
After the first calibration, once a year or after 4,000
hours, whichever comes sooner.
Auto Zero Interval Once per hour (typical)
Leak Tightness < 250 mBar/min when a 30% vacuum is invoked on the flow
system.

Related product manuals