CO2 Specifications
16-8 EarlyVue VS30 Instructions for Use
CO
2
Accuracy 0 mmHg – 38 mmHg: ±2 mmHg
39 mmHg – 150 mmHg: ± (5% of reading + 0.08% for every 1
mmHg above 38 mmHg)
As soon as a CO
2
reading is presented, the measurement is
accurate to the above specifications.
Notes:
Accuracy applies for breath rates up to 80 bpm. For breath
rates above 80 bpm, accuracy is ±4 mmHg or ±12% of reading,
whichever is greater for etCO
2
values greater than 18 mmHg. To
achieve the specified accuracies for breath rates above 60
bpm, use the FilterLine™ H set for infant/neonatal patients.
The accuracy specification is maintained to within 4% of the
values indicated in the presence of interfering gases.
Drift of Measurement
Accuracy
The periodic auto zero function compensates for drifts
between components, changes in ambient temperature, and
barometric conditions. This automatic process eliminates
variances that might otherwise cause measurement drift.
Therefore, the module does not exhibit drift.
Ambient CO
2
Level 0–700 ppm
Respiration Rate
Range
0 rpm to 150 rpm
Respiration Rate
Accuracy
±1 rpm in the range of 0 rpm–70 rpm
±2 rpm in the range of 71 rpm–120 rpm
±3 rpm in the range of 121 rpm–150 rpm
Respiration Rate
Resolution
1 rpm
Automatic Barometric
Pressure
The monitor is equipped with automatic barometric pressure
compensation.
Effects of Cyclical
Pressure
Will operate within specifications with overpressure and
underpressure from a ventilation system as follows:
Overpressure: +100 cmH
2
O
Underpressure: –20 cmH
2
O
IPI Resolution 1
The capnography component of this product is covered by the US patents listed at
www.covidien.com/patents.
1. If the MicroMediCO2 cannot hold the flow within the specified range, the module will stop measuring, enter
Failure Mode, and set the appropriate status bits.