EarlyVue VS30 Instructions for Use 16-9
Temperature Specifications
CO
2
Response Time for Infant/Neonatal Accessories
Temperature Specifications
Predictive Temperature Specifications
Part Number Description
Rise Time
(typical)
Delay
Time
(typical)
Total
System
Response
Time
(typical)
989803159581 VitaLine™ H Set
intubated circuit
240 ms 5.3 s 6.5 s
989803160261 FilterLine™ H Set Long
intubated circuit
240 ms 3.0 s 7.2 s
989803178001 CapnoLine™ H O
2
non-
intubated dual-
purpose circuit
Up to 310 ms Up to 5.9 s 7.2 s
989803178011 CapnoLine™ H Long
non-intubated single-
purpose circuit
Up to 320 ms Up to 5.9 s 7.2 s
989803178021 Nasal FilterLine™ non-
intubated ventilation
single-purpose circuit
Up to 250 ms
(standard
length)
Up to 4.0 s 5.3 s
M1923A FilterLine™ H Set
intubated circuit
180 ms 2.7 s 3.9 s
M4691A CapnoLine™ H non-
intubated single-
purpose circuit
Up to 250 ms Up to 4.0 s 5.3 s
Parameter Specification
Measurement
Range
Monitored (Continuous) Mode: 26.7°C–43.3°C (80.1°F–109.9°F)
Predictive (Spot Check) Mode: 34.4°C – 40.6°C (93.9°F–105.1°F)
Accuracy ±0.1° C or ±0.2° F F in Monitored mode
Measurement
Time
1
Oral: 4 to 6 seconds
Adult axillary (18 years and older): 12 to 15 seconds
Pediatric axillary (under 18 years old): 10 to 13 seconds
Rectal: 10 to 13 seconds
Resolution 0.1°C or ± 0.1° F