Temperature Specifications
16-10 EarlyVue VS30 Instructions for Use
Temporal Temperature Specifications
Heating Transient
Response Time
<150 seconds
Cooling Transient
Response Time
<150 seconds
1. Predictive time depends on probe placement and patient condition. Predictive times do not include tissue
contact detection time or the time needed to place the probe in the measurement site.
Parameter Specification
Parameter Specification
Clinical Accuracy ± 0.1°C or ± 0.2°F per ASTM E1112
Temperature Range 16°C–43°C (61°F–110°F)
Accuracy (overall temperature range) ±0.1° C or ±0.2° F
Arterial Heat Balance Range for Body
Temperature
1
1. Automatically applied when temperature is within normal body temperature range, otherwise reads surface
temperature.
34.5°C–43°C (94°F–110°F)
Normal range is 35.9°C–37.5°C (96.6°F–
99.5°F), with mean of 37°C (98.6°F).
Operating Environment 16°C–40°C (60°F–104°F)
Resolution 0.1°F or 0.1°C on probe or on monitor
Response Time Approximately 1 second
Time Results Displayed on Probe <33 seconds
Size Instrument: 7.9” x 1.8” x 1.6”
Cable: 48” retracted
Weight 0.7 lb
Storage Conditions –20°C to 50°C (–4°F to 122°F)
Display Type and Size Large bright LEDs
Heating Transient Response Time <150 seconds
Cooling Transient Response Time <150 seconds