EarlyVue VS30 Instructions for Use 16-11
SpO
2
Specifications
SpO
2
Specifications
Note
This section describes SpO
2
specifications for a VS30 monitor with Philips SpO
2
. For
information on Masimo SpO
2
specifications, see the EarlyVue VS30 with Masimo
®
Vital
Signs Monitor Instructions for Use.
The update rate for the SpO
2
value and pulse rate is typically less than 1 second. Data
averaging and other signal processing on the displayed and transmitted data values of
SpO
2
and pulse rate is controllable by the user-selectable Sensitivity Mode: Slow
(20 seconds), Normal (10 seconds), and Fast (5 seconds).
Depending on the magnitude of difference between the alarm limit and the displayed
value, the alarm signal generation delay may be from 1 second to the value of the
response time (5, 10, or 20 seconds), plus any additional configured alarm delay time. If
an SpO2 Extd Update technical alarm occurs, the physiological alarm response time is
proportionately longer. If the system is configured for SpO2 Alarm Delay, the alarm
generation is delayed by the configured time.
SpO
2
accuracy is sensor dependent. See the SpO
2
Accuracy section in the table below
and “SpO2 Accuracy Specifications” on page 16-12.
Because pulse oximeter equipment measurements are statistically distributed, only
approximately two-thirds of pulse oximeter equipment measurements can be expected
to fall within the ± Arms value measured by a CO-oximeter.
Parameter Specification
SpO
2
Measurement Range 0%–100%
SpO
2
Measurement Resolution 1%
SpO
2
Accuracy
1
Philips Reusable Sensors Range Accuracy
M1191B, M1191BL, M1192A 70%–100% ± 2%
M1193A, M1194A, M1195A,
M1196A, M1191T, M1192T,
M1196S, M1196T
70%–100% ± 3%
M1193T (Neonatal) 70%–100% ± 4%
Philips Disposable Sensors Range Accuracy
M1131A, M1133A, M1134A
(Neonatal)
70%–100% ± 3%
M1132A, M1133A, M1134A
(Adult/Infant)
70%–100% ± 2%
Pulse Rate Measurement Range 30 bpm–300 bpm
Pulse Rate Accuracy Within 2% or 1 bpm, whichever is greater
2