Specifications
16-14 EarlyVue VS30 Instructions for Use
M1192A
The Philips sensor M1191A was tested as representative of the Philips sensor M1192A
Note
The Philips sensor M1191A is now obsolete and is no longer sold by Philips. It has been
replaced Philips sensor M1191B.
M1193A
The Philips sensor M1193A was tested as representative of the Philips sensor M1195A.
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.40 1.81 1.00 1.22
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.59 2.15 1.33 1.25
SpO
2
- SaO
2
(%SpO
2
)
Upper 95% Limit = +3.08
Mean Difference = +0.48
Lower 95% Limit = -2.11
N = 332, between subject variance = 0.294, within-subject variance = 1.456
BGA Reference (%SaO
2
)
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
Upper 95% Limit = +3.39
Mean Difference = +0.24
Lower 95% Limit = -2.90
N = 333, between-subject variance = 1.153, within-subject variance = 1.424