EarlyVue VS30 Instructions for Use 16-15
SpO
2
Specifications
M1194A
M1196A
The Philips sensor M1196A was tested as representative of the Philips sensors M1196S
and M1196T.
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.42 1.39 1.41 1.44
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.90 2.82 1.33 1.35
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
N = 330, between-subject variance = 1.329, within-subject variance = 0.686
Upper 95% Limit = +2.44
Mean Difference = -0.34
Lower 95% Limit = -3.12
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
Upper 95% Limit = +3.10
Mean Difference = -0.52
Lower 95% Limit = -4.14
N = 327, between-subject variance = 1.001, within-subject variance = 2.414