SpO
2
Specifications
16-16 EarlyVue VS30 Instructions for Use
M1191T
The Philips sensor M1191T was tested as representative of the Philips sensor M1192T.
M1193T
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.50 2.13 1.15 0.99
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.80 2.72 1.46 0.98
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
Upper 95% Limit = +3.09
Mean Difference = +0.13
Lower 95% Limit = -2.82
N = 332, between-subject variance = 0.648, within-subject variance = 1.630
N = 333, between-subject variance = 1.775, within-subject variance = 1.631
BGA Reference (%SaO
2
)
SpO
2
- SaO
2
(%SpO
2
)
Upper 95% Limit = +3.70
Mean Difference = +0.08
Lower 95% Limit = -3.54