EarlyVue VS30 Instructions for Use 16-17
SpO
2
Specifications
M1131A
.
M1133A
The Philips sensor M1133A was tested as representative of the Philips sensors M1132A
and M1134A.
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.81 2.53 1.57 1.29
SaO
2
range (%) 70-100 70-80 80-90 90-100
A
rms
value (%) 1.50 1.76 1.61 1.16
N = 324, between-subject variance = 1.634, within-subject variance = 1.736
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
Upper 95% Limit = +3.38
Mean Difference = -0.22
Lower 95% Limit = -3.82
N = 331, between-subject variance = 0.978, within-subject variance = 0.895
SpO
2
- SaO
2
(%SpO
2
)
BGA Reference (%SaO
2
)
Upper 95% Limit = +3.36
Mean Difference = +0.68
Lower 95% Limit = -2.01