EarlyVue VS30 Instructions for Use A-5
Technical Alarms
Battery Overheat,
shutting down
High Battery has overheated (exceeded 60°C).
Note: Charging stops immediately. The monitor shuts
down when the message is acknowledged or after 5
seconds if not acknowledged.
Replace Battery Low Battery Cycle Count value exceeds 300 or the
battery SOH is less than or equal to 80%.
Replace Battery
Immediately
High Battery Cycle Count value exceeds 400 or the
battery SOH is less than or equal to 75%.
High Battery Temp,
charge stopped
Low The battery temperature exceeds 50°C.
Battery Overheat High Battery temperature exceeds high limit of 60°C.
Philips Device Management Dashboard Messages
Configuration
Update Failed
Low The monitor was unable to import settings from the
Philips Device Management Dashboard server. This
is only a visual message.
Configuration
Update Requested
Low The Philips Device Management Dashboard server
sends a request to import configuration settings to
the monitor. This is only a visual message.
Software Upgrade
Failed
Low The Philips Device Management Dashboard server
indicates that the software upgrade was not
successful. This is only a visual message.
Software Upgrade
Requested
Low The Philips Device Management Dashboard server
sends a request to upgrade the monitor software.
This is only a visual message.
NBP Alarms
NBP Air Leak Low The monitor cannot adjust pressure. This may be
due to leakage or an internal NBP module problem.
NBP Artifact Low The monitor cannot correct the pressure to the
intended value within the time limit, or the monitor
requires too many pressure correction attempts to
adjust the pressure to the intended value.
This may be due to excessive patient movement,
leakage, or a problem with extreme edematous
patients.
NBP Equip Malfunc Low NBP equipment malfunction. See your system
administrator to check the error log for details.
NBP Hose Blocked Low The monitor has detected a defect in the
pneumatic system, such as valves, tubing, or plug.
NBP Loose Cuff Low The NBP cuff cannot inflate to the target value
within the limits of the selected patient size. May
be caused by a pump defect, leakage, or
disconnected cuff.
Alarm Message Priority Cause