EarlyVue VS30 Instructions for Use A-7
Technical Alarms
Temperature Alarms
Temp Module
Malfunc
Low There is a temperature module malfunction.
Predictive Temperature
The malfunction may be caused by electrostatic
discharge on the temperature probe. Reset the
temperature module and clear the error by
inserting the probe in the probe well and pulling it
out. If the error message is not cleared, the
malfunction may be caused by one of the
following:
Battery or power supply voltage not in range
Ambient temperature too low or too high
See your system administrator to check the error
log for details.
Temporal Temperature
The malfunction may be caused by one of the
following:
Ambient temperature too low or too high
Checksum error
Internal firmware error
See your system administrator to check the error
log for details.
Temp Out of Range Low The temperature value is outside the temperature
measurement range.
Temp Probe Error Low Predictive Temperature
The malfunction may be caused by electrostatic
discharge on the temperature probe. Reset the
temperature module and clear the error by
inserting the probe in the probe well and pulling it
out. If the error message is not cleared, the
malfunction may be caused by one of the
following:
The probe well is missing
The probe well is not installed properly
The probe warmer has overheated
See your system administrator to check the error
log for details.
Alarm Message Priority Cause