Factory Default Alarm Limits and Alarm Ranges
A-8 EarlyVue VS30 Instructions for Use
Factory Default Alarm Limits and Alarm Ranges
This section lists the default alarm limits and the alarm limit ranges for all physiological
alarms.
Factory Default Alarm Limits
The following table lists the default alarm limits that are set in the factory.
Note
Your system administrator can change these factory defaults to different default values.
Adult Pediatric Neonatal
High Low High Low High Low
etCO2 60 mmHg
(8.0 kPa)
25 mmHg
(3.3 kPa)
60 mmHg
(8.0 kPa)
25 mmHg
(3.3 kPa)
60 mmHg
(8.0 kPa)
25 mmHg
(3.3 kPa)
imCO2 4 mmHg
(0.5 kPa)
NA 4 mmHg
(0.5 kPa)
NA 4 mmHg
(0.5 kPa)
NA
IPI NA 5 NA NA NA NA
NBP Diastolic 90 mmHg
(12.0 kPa)
50 mmHg
(6.7 kPa)
70 mmHg
(9.3 kPa)
40 mmHg
(5.3 kPa)
60 mmHg
(8.0 kPa)
20 mmHg
(2.7 kPa)
NBP Mean 110 mmHg
(14.7 kPa)
70 mmHg
(9.3 kPa)
90 mmHg
(12.0 kPa)
50 mmHg
(6.7 kPa)
70 mmHg
(9.3 kPa)
24 mmHg
(3.2 kPa)
NBP Systolic 160 mmHg
(21.3 kPa)
90 mmHg
(12.0 kPa)
120 mmHg
(16.0 kPa)
70 mmHg
(9.3 kPa)
90 mmHg
(12.0 kPa)
40 mmHg
(5.3 kPa)
Pulse Rate 120 bpm 50 bpm 160 bpm 75 bpm 200 bpm 100 bpm
Respiration
(CO2)
30 rpm 8 rpm 30 rpm 8 rpm 100 rpm 30 rpm
SpO
2
100% 90% 100% 90% 95% 85%
Temperature 39°C
(102.2°F)
36°C
(96.8°F)
39°C
(102.2°F)
36°C
(96.8°F)
39°C
(102.2°F)
36°C
(96.8°F)