EarlyVue VS30 Instructions for Use A-9
Factory Default Alarm Limits and Alarm Ranges
Alarm Limit Ranges
The following table lists the user-adjustable ranges for all physiological alarms.
Adult Pediatric Neonatal
etCO
2
High 32 mmHg – 149 mmHg
(4.3 kPa – 19.9 kPa)
32 mmHg – 149 mmHg
(4.3 kPa – 19.9 kPa)
32 mmHg – 149 mmHg
(4.3 kPa – 19.9 kPa)
etCO
2
Low 1 mmHg – 70 mmHg
(0.1 kPa – 9.3 kPa)
1 mmHg – 75 mmHg
(0.1 kPa – 10 kPa)
1 mmHg – 75 mmHg
(0.1 kPa – 10 kPa)
imCO
2
High 2 mmHg – 20 mmHg
(0.3 kPa – 2.7 kPa)
2 mmHg – 20 mmHg
(0.3 kPa – 2.7 kPa)
2 mmHg – 20 mmHg
(0.3 kPa – 2.7 kPa)
imCO2 Low NA NA NA
IPI 1 – 10 NA NA
NBP Diastolic
High
55 mmHg – 239 mmHg
(7.3 kPa – 31.9 kPa)
55 mmHg – 149 mmHg
(7.3 kPa – 19.9 kPa)
22 mmHg – 99 mmHg
(2.9 kPa – 13.2 kPa)
NBP Diastolic
Low
11 mmHg – 85 mmHg
(1.5 kPa – 11.3 kPa)
11 mmHg – 65 mmHg
(1.5 kPa – 8.7 kPa)
11 mmHg – 55 mmHg
(1.5 kPa – 7.3 kPa)
NBP Mean
High
65 mmHg – 249 mmHg
(8.7 kPa – 33.2 kPa)
55 mmHg – 159 mmHg
(7.3 kPa – 21.2 kPa)
26 mmHg – 119 mmHg
(3.5 kPa – 15.9 kPa)
NBP Mean
Low
21 mmHg – 105 mmHg
(2.8 kPa – 14.0 kPa)
21 mmHg – 85 mmHg
(2.8 kPa – 11.3 kPa)
21 mmHg – 65 mmHg
(2.8 kPa – 8.7 kPa)
NBP Systolic
High
95 mmHg – 269 mmHg
(12.7 kPa – 35.9 kPa)
75 mmHg – 179 mmHg
(10.0 kPa – 23.9 kPa)
45 mmHg – 129 mmHg
(6.0 kPa – 17.2 kPa)
NBP Systolic
Low
31 mmHg – 155 mmHg
(4.1 kPa – 20.7 kPa)
31 mmHg – 120 mmHg
(4.1 kPa – 16.0 kPa)
31 mmHg – 85 mmHg
(4.1 kPa – 11.3 kPa)
Predictive
Temperature
High
36.1°C – 43.2°C
(97.0°F – 109.8°F)
36.1°C – 43.2°C
(97.0°F – 109.8°F)
36.1°C – 43.2°C
(97.0°F – 109.8°F)
Predictive
Temperature
Low
26.8°C – 38.9°C
(80.2°F – 102.0°F)
26.8°C – 38.9°C
(80.2°F – 102.0°F)
26.8°C – 38.9°C
(80.2°F – 102.0°F)
Pulse Rate
High (SpO2)
55 bpm – 299 bpm 80 bpm – 299 bpm 80 bpm – 299 bpm
Pulse Rate
Low (SpO2)
31 bpm – 115 bpm 31 bpm – 155 bpm 31 bpm – 195 bpm
Pulse Rate
High (NBP)
55 bpm – 299 bpm 80 bpm – 299 bpm 80 bpm – 299 bpm
Pulse Rate
Low (NBP)
41 bpm – 115 bpm 41 bpm – 155 bpm 41 bpm – 195 bpm
SpO2 High 50% – 100% 50% – 100% 31% – 100%