EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
46
Export 3D Volume Data
Export Temporal (4D) Volume
Export 3D Volume Presentation State
Click Save.
Hard Drive Status and Maintenance Settings
When the hard drive capacity is close to being reached the hard-drive
status messaging feature informs users of the remaining hard drive
capacity. The hard drive status messages prompt users to back up and
delete the patient database before capacity is reached, to reduce the
incidence of database corruption, minimize system downtime and
prevent loss of patient data. These warning messages will appear
when the user touches Patient or Review:
Additional images can still be saved and additional patient exams can
be created for Almost full” messages.
Full” messages do not allow additional patients to be created or
images to be saved. If the maximum number of patients has been
entered, the patient data fields in the patient data form are
unavailable.
The following conditions trigger these status messages:
There is less than 15% drive space or there is insufficient space
for ten exams remaining.
The maximum space for images and loops (multi-frame objects)
has been reached.
The maximum space for exams has been reached.
4535 616 21547 EPIQ Series Field Service Manual Page 221
To specify the auto-delete exam capabilities for maintaining the hard
drive:
Press Support and click the Test/Utilities tab.
Click Disk Utilities Configure Auto-Delete.