EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
56
To Test connectivity and review functionality:
1. Start an exam.
2. Press 2D and press Color.
3. Turn Color to adjust color gain, until color noise appears in the
color box.
4. Press the Acquire control that is configured to a printer. Verify
that you hear a beep, that a print is made (if an Acquire control is
mapped to a specific printer), and that the study thumbnail
appears to the right of the image.
5. Touch Patient.
6. Touch End Exam and click Done.
7. To display the Patient Directory, touch Review.
8. Insert a new DVD into the DVD drive.
9. Select the patient exam that was most recently ended and click
Send To.
10. In the Send To dialog box, click the Media tab, select DVD, and
click OK.
11. Verify that the DVD status is Busy.
12. Delete the patient file that was copied to the DVD from the hard
drive.
13. For the Source, select DVD/CD.
14. Select the patient file you copied to the DVD, and click Send to
HD.
15. Select the hard drive directory and the exam that was copied from
the DVD. Click Open.
16. Double-click the JPEG loop icon to obtain a full-screen playback of
that loop.
17. Click Done.