EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
62
Click System Test Run.
Click Yes. The system test results are displayed in the System
Test box.
Click OK.
To export the system test results:
Perform the System Test.
In the System Test Results area, click Export Logs.
Connect a USB flash drive to a USB connector and select that port
as your export destination.
In Select Logs to Export, click Include Diagnostic Logs. You may
export other logs also, but the system test results are displayed
in the Diagnostic Logs.
Check the Status area for a successful export and click OK.
Filter Cleaning
Philips recommends once yearly cleaning at minimum and more often
in dusty environment. Also, you should think of this as one of these
very first “Is it plugged in?” troubleshooting steps. The system shuts
down if it overheats. The first and simplest probable cause is lack of
air flow.
Filter Cleaning Procedure
System power must be turned off.
Locate and slide filter out of left side.
Replace with spare filter and clean the filter first by vacuum,
wash with soap & water and dry. If a spare is not available
remember that the filter must be dry before replacement. A
damp filter could damage the system!
Reset the system’s “filter last cleaned date as outlined below.