Set Filter Status Procedure
Press “Support” and log on using IST
Click: Test/Utilities Service Messaging Check Filter Status
Enter the “Next Filter Cleaning date” and click on “Interval Days
to enter the interval length.
Clicking “Done” sets the date.
You should verify that the “Next Filter Cleaning Date” has been
updated appropriately.
System Lift Tool and Servicing Casters
The recommended interval for caster maintenance is also every 12
months. You will need to use the system cart lift tool and Molygraph
SL3144 bearing grease as recommended by Philips if maintence or
replacement is needed.
Move the system, check caster operation and visually inspect for
damage.
Only the front casters have brake/steering locks. The back casters
swivel for steering. The single pedal at the front of the system
controls the brake and steering. The pedal has three positions:
Fully up The casters are steering locked; the system rolls freely
but does not steer.