7 Patient Alarms and INOPs
145
CL SpO2 Pod INOPs
SpO₂ Upgrade Monitor: display only
Numeric unavailable
The SpO
2
measurement is currently in upgrade mode.
Monitoring is not possible in this mode.
SpO₂Interference Monitor: INOP tone
Numeric is replaced by a -?-
There is too much interference caused by a high level of
ambient light, and/or electrical interference. Cover the
sensor to minimize ambient light. If the INOP persists,
ensure that the sensor cable is not damaged, or positioned
too close to power cables.
INOP Message Indication What to do
INOP Message Indication What to do
Check Charger I/F Monitor: INOP tone
CL SpO
2
Pod: display
message
Overvoltage or undervoltage detected at the charger
interface. Clean contacts of charger interface at Cableless
Measurement Device and charging station. If the INOP
persists, contact your service personnel.
cl SpO₂ Batt Empty
!!cl SpO₂ BatEmpty
!!!cl SpO₂ BtEmpty
Monitor: INOP tone
(configurable cyan, yellow,
red)
CL SpO
2
Pod: display
message, yellow tone
The remaining battery time of the CL SpO
2
Pod is
below 30 minutes. Charge the battery.
cl SpO₂ Batt Incmp Monitor: INOP tone
CL SpO
2
Pod: display
message
The battery in use with the CL SpO
2
Pod is incompatible.
Replace it with one approved for use with the CL SpO
2
Pod.
cl SpO₂ Batt Low Monitor: INOP tone
CL SpO
2
Pod: display
message
The remaining battery time of the CL SpO
2
Pod is below
2 hours.
cl SpO₂ Batt Malf Monitor: INOP tone
CL SpO
2
Pod: display
message
There is a malfunction in the CL SpO
2
Pod's battery
system. Contact your service personnel.
cl SpO₂ Batt Temp Monitor: INOP tone
CL SpO
2
Pod: display
message
The temperature of the battery in the CL SpO
2
Pod is
critically high. Check that the Pod is not covered or
exposed to a heat source. If the INOP persists, remove
the CL Pod from the patient, and contact your service
personnel.
cl SpO₂ Check Batt Monitor: INOP tone
CL SpO
2
Pod: display
message
The battery in the CL SpO
2
Pod is nearing the end of its
useful life. Only 50 charge-discharge cycles remain.
Contact your service personnel to replace the battery.