3 Basic Operation
72
Monitoring After a Power Failure
A fetal monitor that was switched on prior to a temporary power loss, switches on again when
power is restored.
A fetal monitor that was switched off prior to a temporary power loss, remains off when power is
restored.
If the fetal monitor is without power for less than one minute, monitoring will resume with all
active settings unchanged.
If the fetal monitor is without power for more than one minute, the behavior depends on your
configuration.
–If
Automat. Default is set to Yes, the default profile will be loaded when power is restored.
–If
Automat. Default is set to No, all active settings are retained, if power is restored within
48 hours. The
Automat. Default setting is made in Configuration Mode.
Troubleshooting
Problem Possible Causes Solutions
Light or no trace Wrong paper Use recommended paper
Dirty printhead Clean printhead, see “Cleaning the
Print Head” on page 273
FM20/30 only: Paper misaligned due
to drawer not being correctly shut
Shut the drawer fully, pushing evenly
with both hands
End of paper noted when pack not
finished
Bad paper feed or wrong paper Check paper feed and use
recommended paper
Check Paper INOP is displayed INOP messages always indicate
equipment problems
See “Patient Alarms and INOPs” on
page 129
FetRec Equip Malf INOP is displayed
Paper End INOP is displayed
Wrong Paper Scale INOP is displayed