3-2
PHILIPS MEDICAL SYSTEMS
STEP 1: PREPARATION
Press the On/Off button to turn on the HeartStart FR2+
Defibrillator. Follow the instructions provided by the FR2+ voice
and text prompts.
Remove clothing from the patient's upper body. If needed,
wipe moisture from the patient's skin and clip or shave excessive chest hair.
If the patient appears to be under eight years of age or 55 lb. (25 Kg), use
M3870A FR2 infant/child reduced-energy defibrillator pads, if available.
If the infant/child pads are not available, or if the child appears older/larger,
use adult defibrillator pads.
DO NOT DELAY TREATMENT TO
DETERMINE THE CHILD’S EXACT AGE/WEIGHT.
Open the defibrillator pads package. Pull off the protective backing from the
defibrillator pads.
Place the sticky side of each pad on the patient’s bare skin, exactly as shown
on the drawing on each pad.
Insert the defibrillator pads connector firmly into the
defibrillator’s connector socket, indicated by a flashing
light at the top left of the FR2+.
O
N
O
F
F
Pad placement for adults and children 8 years old
or 55 pounds or more.
Pad placement for infants and children
under 8 years old or 55 pounds.