PHILIPS MEDICAL SYSTEMS
5
DEFIBRILLATION WARNINGS AND CAUTIONS
CAUTION Alteration of the factory default setup of the FR2+ can affect its performance and
should be performed under the authorization of your Medical Director. Modifications
to device operation resulting from changes to the default settings should be specifically
covered in user training.
CAUTION Use only Philips-approved data cards. The HeartStart FR2+ may record improperly if a
non-approved data card is used. Install empty data card after each use to avoid loss of
data.
safety level possible electrical interference with ECG monitoring
WARNING Radio-frequency (RF) interference from devices such as cellular phones and two-way
radios can cause improper HeartStart FR2+ operation. The HeartStart FR2+ should
be used at least 6 feet (2 meters) away from RF devices, as stated in accordance with
EN 61000-4-3:2002.
safety level possible improper device performance
safety level possible shock hazard
WARNING Defibrillation current can cause operator or bystander injury. Do not touch the
patient during defibrillation. Do not allow the defibrillator pads to touch any metal
surfaces. Disconnect the pads connector from the HeartStart FR2+ before using any
other defibrillator.
WARNING Before delivering a shock, it is important to disconnect the patient from other medical
electrical equipment, such as blood-flow meters, that may not incorporate
defibrillation protections. In addition, make sure the pads are not in contact with metal
objects such as a bedframe or stretcher.
safety levels possible ECG misinterpretation
WARNING For patient safety reasons, some very low-amplitude or low-frequency heart rhythms
may not be interpreted by the HeartStart FR2+ as shockable VF rhythms. Also, some
VT rhythms may not be interpreted as shockable rhythms.
WARNING Handling or transporting the patient during heart rhythm analysis can cause an
incorrect or delayed diagnosis. If the HeartStart FR2+ gives a
SHOCK ADVISED
prompt, keep the patient as still as possible for at least 15 seconds so the HeartStart
FR2+ can reconfirm the rhythm analysis before a shock is delivered.