PHILIPS MEDICAL SYSTEMS
B-1
B
B TECHNICAL SPECIFICATIONS
The specifications for the HeartStart FR2+ provided in this chapter apply to
both the M3860A and M3861A, unless otherwise noted. Additional information
can be found in the Technical Reference Manuals for HeartStart Defibrillators,
located online at www.philips.com/productdocs.
HEARTSTART FR2+ DEFIBRILLATOR SPECIFICATIONS
PHYSICAL
ENVIRONMENTAL
category nominal specifications
size 2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).
weight 4.7 lbs (2.1 kg) with M3863A FR2 standard battery installed.
4.5 lbs (2 kg) with optional M3848A FR2+ rechargeable battery installed.
category nominal specifications
temperature and relative
humidity
Operating (battery installed, pads connected): 32° to 122° F (0° to 50° C);
0% to 95% relative humidity (non-condensing).
Standby (battery installed and stored with defibrillator pads): 32° to 109° F
(0° to 43° C); 0% to 75% relative humidity (non-condensing).
Storage/shipping: -4° to 140° F (-20° to 60° C); 85% relative humidity
(non-condensing)
altitude Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).
shock/drop abuse
tolerance
Meets MIL-810E 516.4, Procedure IV (after a 1 meter drop to any edge,
corner, or surface, in standby mode).
vibration Meets MIL-810E 514.4-16 and 514.4-17.
sealing With data card tray and battery installed, meets IEC 529 class IP54.
ESD/EMI See Appendix F.
aircraft: method Meets RTCA/DO-160D:1997 Section 21 (Category M - Charging).