PHILIPS MEDICAL SYSTEMS
F-2
ELECTROMAGNETIC EMISSIONS
ELECTROMAGNETIC IMMUNITY
emissions test compliance electromagnetic environment – guidance
RF CISPR 11 Group
1
Class B
The FR2+ uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The FR2+ is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
immunity test
IEC 60601
test level
compliance level
electromagnetic environment -
guidance
electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
There are no special requirements
with respect to electrostatic
discharge.
a
power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial/hospital environment.
There are no special requirements for
non-commercial/non-hospital
environments.
radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.5 GHz
20 V/m Portable and mobile RF
communications equipment should be
used no closer to any part of the
HeartStart FR2+, including cables,
than is absolutely necessary.
b,c
The
recommended separation distances
for various transmitters and the AED
are shown in the following table.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.