EasyManua.ls Logo

Philips HeartStart OnSite - Technical Specifications and Waveforms

Philips HeartStart OnSite
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PHILIPS MEDICAL SYSTEMS
E-1
E
E TECHNICAL INFORMATION
HEARTSTART ONSITE DEFIBRILLATOR SPECIFICATIONS
The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manuals for HeartStart
Automated External Defibrillators, located online at www.philips.com/
productdocs.
PHYSICAL
ENVIRONMENTAL
category specifications
size 7.2cm H x 19cm D x 21cm W (2.85” H x 7.40” D x 8.30” W).
weight Approximately 1.5 kg (3.3 lbs) with battery and pads cartridge installed.
category
specifications
temperature and
relative humidity
Operating (battery and pads cartridge installed):
0° to 50° C (32° to 122° F);
0% to 95% RH (non-condensing).
Standby (between uses with battery and pads cartridge installed):
10° to 43° C (50° to 109° F);
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads cartridge):
-2to 60° C (-4° to 140° F) for up to 2 days;
0% to 85% RH (non-condensing).
altitude Operates at 0 to 4,572 m (15,000 feet); can be stored at up to 2,591 m
(8,500 feet), in standby mode.
atmospheric pressure Operates at 1013 hPa to 590 hPa; can be stored at up to 750 hPa, in standby
mode.
shock/drop abuse
tolerance
Withstands 1 meter (3.3 foot) drop to any edge, corner, or surface.

Other manuals for Philips HeartStart OnSite

Related product manuals