PHILIPS MEDICAL SYSTEMS
F-2
PATIENT TREATMENT PROTOCOL OPTIONS
parameter settings default default description
“call EMS” voice
reminder timing
At power on (when
the user turns on
the OnSite)
At power on and at
the start of the first
patient care pause
At the start of the
first patient care
pause
No reminder
At the start of
the first patient
care pause
Provides a voice reminder to make
sure emergency medical services
have been called, at the start of the
first patient care pause.
shock series 1, 2, 3, 4 1 The automatic protocol pause for
CPR is activated each time a shock is
delivered.
*
During the protocol pause, the
OnSite does not perform rhythm
analysis.
The length of the protocol pause
after a shock series is completed is
determined by the protocol pause
timer setting.
shock series interval
(minutes)
1.0, 2.0,
f(infinity)
1.0 A delivered shock must occur within
1 minute of the previous shock to be
counted as part of the current shock
series.
NOTE: This parameter is only
applicable when the shock series is not
configured to the default 1 shock.
* A shock series begins when a shock is delivered after the OnSite is turned on. A new shock series begins after a
protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the
previous shock exceeds the shock series interval setting.