Technical specifications
Mains power supply Input = 100 – 240 V~, 50/60 Hz
Output = 5V
, 1.0A
Internal rechargeable battery 3.7 Volts nominal, 1200 mAh
power supply (lithium polymer)
Weight 0.29 lbs/111g
Size 7.0cm x 4.5cm x 13.5cm
Class II device internally powered device (double safety insulation)
Type BF device (device with specific protection against electrical hazards)
Ingress Protection rating IP22. (Protected against solid foreign objects of 12.5mm diameter and greater;
protected against vertically falling water drops when device titled upto 15°) The InnoSpire Go handset and
charger form a Medical Electrical System, in which the charger is not Medical Electrical Equipment. The IP22
rating applies to the InnoSpire Go handset, the charger rating is IPX0.
All components shown in the illustration are applied parts.
Materials: Handset - Polyamide (PA) and Thermoplastic elastomer (TPE),
Mouthpiece Assembly - Polyamide (PA) and Polypropylene (PP)
Certification
Reference to standards
Electric safety standards EN 60601-1
Electromagnetic compatibility according to EN 60601-1-2
Operating conditions
Temperature range of + 5 °C to + 40 °C
Humidity range of 15% RH to 93% RH, non-condensing
Atmospheric pressure 70 kPa to 106 kPa
Storage and transport conditions
Temperature range of MIN -25° C MAX +70° C
Humidity range of MIN 10% RH –MAX 93% RH
Atmospheric pressure 50 kPa to 106 kPa
Replacement parts and optional accessories
Adult mask - large (ages 5 years plus) 1127875
Pediatric mask - medium (ages 1 – 5 years) 1127798
Pediatric mask - small (ages 0 – 18 months) 1127822
Mask adapter 1125985
Mouthpiece assembly 1128501
European (2 pin) plug adapter 1127650
UK (3 pin) plug adapter 1127652
Carry case 1128576
Battery has insufficient charge, follow battery charging
instructions.
If the problem persists, contact your customer services
representative.
Troubleshooting
Problems Solutions
The on button is pressed but
nothing happens (no light, no
nebulization).
LED illuminates on button
press but goes off when
button released.
Check to ensure that the mouthpiece assembly is attached
properly.
Make sure that the contacts on the handset are dry and free
of debris.
Make sure that there is medication loaded in the chamber.
Recharge the battery.
Treatments take longer
than usual with the same
mouthpiece assembly.
Should the device still not operate properly after checking the unit as indicated above,
contact Philips
Customer Service at 1-800-345-6443 or 1-724-387-4000.
Boil the mouthpiece assembly as per the weekly
disinfection instructions.
When charging the battery,
the LED does not pulse or
turn on.
Ensure that the power cable is connected to the handset
and plugged into the wall outlet.
Ensure nebulizer is held upright during nebulization.
If the problem persists, contact your customer services
representative.
Device indicates end of
treatment but medication
remains in medication chamber
(more than a few drops).
Device does not indicate end
of treatment even though all
medication is nebulized.
Contact your customer services representative.