EasyManua.ls Logo

Philips REMstar Pro C-Flex+ - Page 14

Philips REMstar Pro C-Flex+
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 User Manual
Status - This displays information sent from a peripheral (SD card , modem , etc.). If two peripherals are
attached, two lines will appear with corresponding icons.
Note: This will not display if no peripherals are being used.
Phone-in - This screen displays the total therapy hours for the device, the total blower hours, and the total
number of days used when the sessions were greater than 4 hours since the device was last reset by the home care
provider. This screen also displays a compliance check number used by your home care provider to validate that
the data provided by you is the data taken from this screen. This setting only appears if your provider has enabled
this feature.
Compliance VIC (Visual Inspection Check) - This screen displays the start day and the total number of days
used when the sessions were greater than 4 hours. This screen also displays a check code number used by your
home care provider to validate that the data provided by you is the data taken from this screen. This setting only
appears if your provider has enabled this feature.
Therapy hours - The device is capable of recognizing the difference between the time the patient is actually
receiving therapy and the time when the blower is simply running. This screen displays the average amount of time
the patient is actually receiving therapy on the device over a 7 day and 30 day time frame (provided the device has
at least 7 or 30 days of data respectively). If the device has only 5 days of data to use for the calculation, the 5 day
average value will be seen under the 7 day display.
Days > 4 - This screen displays the cumulative number of device therapy sessions that exceeded 4 hours over a 7
day and 30 day time frame.
Large leak - During any given night, the device recognizes the percentage of time the patient was experiencing
what it deemed to be a large leak. Large leak is dened as the level of leak that is so large, it is no longer possible
to determine respiratory events with statistical accuracy. This screen displays the average of these individual
nightly values of percentage of time in large leak over a 7 day and 30 day time frame (provided the device has
at least 7 or 30 days of data respectively). If the device has only 5 days of data to use for the calculation, the 5
day average value will be seen under the 7 day display. If you see a large increase in the percent of time in large
leak indicated here, contact your home care provider for assistance. This screen only displays if your home care
provider has enabled it.
AHI - The device accumulates individual Apnea/Hypopnea indices (AHI) for each session the patient used the
device. This screen displays the average of these individual nightly AHI values over a 7 day and 30 day time frame
(provided the device has at least 7 or 30 days of data respectively). If the device has only 5 days of data to use for
the calculation, the 5 day average value will be seen under the 7 day display. This screen only displays if your home
care provider has enabled it.
Periodic Breathing - During any given night, the device recognizes the percentage of time the patient was
experiencing periodic breathing. This screen displays the average of these individual nightly values of periodic
breathing over a 7 day and 30 day time frame (provided the device has at least 7 or 30 days of data respectively). If
the device has only 5 days of data to use for the calculation, the 5 day average value will be seen under the 7 day
display. If you see a large increase in the percent of time in periodic breathing indicated here, contact your home
care provider for assistance. This screen only displays if your home care provider has enabled it.
Auto-Trial - If the device is in the Auto-Trial mode, this screen displays Days: 00/00 (accumulated trial days /
selected trial days). This screen only displays if your home care provider has enabled Auto-Trial.
CPAP-Check - If the device is in the CPAP-Check mode, this screen will either display 00.0 (CPAP-Check
Pressure) or 90%(00.0) (if Auto-Trial established a 90% pressure level). This screen will also display 00/30 (Hours
Used / 30 Hours). This screen only displays if your home care provider has enabled CPAP-Check.
Humidier - This screen will display 3 settings: power supply (either the 60W or 80W), tubing type, and either
humidier or tube temperature setting (if using).