11
EverFlo / EverFlo Q User Manual
English
Chapter 5: Specifications
Environmental
Operating Transport & Storage
Temperature 13 to 32° C -34 to 71° C
Relative Humidity 15 to 95%, noncondensing 15 to 95%, noncondensing
Altitude 0 to 2286 m N/A
Physical
Dimensions 58 cm x 38 cm x 24 cm
Weight 14 to 15 kg
Standards Compliance
This device is designed to conform to the following standards:
- IEC 60601-1 Medical Electrical Equipment, Part 1: General Requirement for Safety
- IEC 60601-1-2 2nd edition, Medical Electrical Equipment, Part 1-2: General Requirement for Safety - Collateral Standard: Electromagnetic
Compatibility - Requirements and Tests
- ISO 8359 Oxygen Concentrators for Medical Use - Safety Requirements
Electrical, AC Power Consumption
Models 1020000, 1020001
1020002, 1020003
1020002BR, 1020003BR
1020014, 1020015
1039362, 1039363
1039364, 1039365
120 VAC ±10%, 350 W, 60 Hz
1020004, 1020005 230 VAC ±10%, 320 W, 60 Hz
1020006, 1020007, 1020008
1020009, 1020010
1020011, 1020012
1020016, 1020017
1020020, 1039366
1039367, 1039368
1039370, 1104000
230 VAC ±10%, <300 W, 50 Hz
1020013, 1102443
230 VAC ±10%, <300 W, 60 Hz