Manufacturer’s Declaration
ThepM-K05 JUMBO is intended for use in the electromagnetic environment
specied below. The customer or the user of thepM-K05 JUMBO should assure
that it is used in such an environment.

Emission Test Compliance Electromagnetic Environment
RF emission CISPR 11 Group 1
ThepM-K05 JUMBO uses RF en-
ergy only for internal functions.
Therefore, this RF emission is
extremely weak and there is little
chance of it creating any kind of in-
terference whatsoever with nearby
electronic equipment.
RF emissions CISPR 11 Class B ThepM-K05 JUMBO is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public low
voltage power supply network that
supplies buildings used for domes-
tic purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage uctuations/icker IEC
61000-3-3
Not applicable

Immunity test IEC60601-1-2 test level
Compliance
level
Electromagnetic environ-
ment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electric fast
transient/ burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV differential mode
±2 kV common mode
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.