110-240 15
Setting Operational Parameters
The Chiller’s various operational parameters, such as temperature, flow rate, and pressure alarm values, are all
user-adjustable. They are accessed by pressing and holding the Units/Menu Button until HL appears on the
pressure/flow rate display. Pressing and releasing the Units/Menu Button once HL appears allows you to scroll
through the various parameters; rotating the Select/Set knob allows you to change the displayed setting. You can
accept the displayed value by either pressing the Select/Set knob or allowing the display to timeout.
Menu Item Description
Reset by Unit
Change
(°C -°F)
Choice/Ranges / Comments
Default
Setting
HL
High Temperature Limit
Alarm Set Point
Yes +15°C to 71°C / 58° to 159°F
71ºC
159°F
LL
Low Temperature Limit
Alarm Set Point
Yes -46°C to +14°C / -50° to 58°F
-46ºC
-50°F
HA
High Ambient Temperature
Alarm Set Point
Yes
+30° to 71°C.
Always displayed and set in °C.
50ºC
FP w/psi
LED lit
Maximum Fluid Pressure
Alarm Set Point
Yes 40 to 100 PSI 100 PSI
FP w/kPa
LED lit
Maximum Fluid Pressure
Alarm Set Point
Yes 280 to 690 kPa 680 kPa
FL w/Option E
LED lit
Minimum Flow Rate Alarm
Set Point
Yes 0 or 0.8 to 2.0 GPM
0.0
GPM
FL w/Option
M LED lit
Minimum Flow Rate Alarm
Set Point
Yes 0 or 3 to 8 LPM
0.0
LPM
Sd
Establishes the cooling
and heating rate by setting
a desired differential
between the measured
internal and external
temperature.
Yes
4°C to 20°C
Always displayed and set in °C
NOTE: Sd is displayed only if the optional
remote temperature control probe is
installed.
5°C
AF
Auto-Refrigeration
Temperature Set Point
No
+20° to 50°C
Always displayed/set in °C.
40ºC
rP
External temperature probe
status
No
n-A (the remote probe is not available)
nEP (the remote probe is not plugged in)
NO (the remote probe is disabled)
AtC (ambient temperature control probe
enabled)
rPC (remote temperature control probe
enabled)
n-A
c
1
Internal Probe Calibration
Offset
No
±2.9°C.
Always displayed/set in °C.
Special access procedure required.
See Internal Calibration Offset
0.0ºC
c
2
External Probe Calibration
Offset (only appears if a
second probe is connected)
No
±2.9°C.
Always displayed/set in °C.
Special access procedure required.
See “External Calibration Offset
0.0ºC
PC RS232 Baud Rate No
00, 24, 48, 96, 192.
Represents baud rates of 0 (no
communication), 2400, 4800, 9600, and
19200.
96

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PolyScience Chiller and is the answer not in the manual?

PolyScience Chiller Specifications

General IconGeneral
BrandPolyScience
ModelChiller
CategoryChiller
LanguageEnglish

Related product manuals