Operator’s Manual
LS-Series Compact Chiller
110-445 12.20.12

Questions and Answers

  • J
    jill66Jul 31, 2025
    Why is my PolyScience Chiller not cooling properly?
    • J
      Jason TaylorJul 31, 2025
      If your PolyScience Chiller isn't cooling effectively, there might be a few reasons. First, check for dust buildup on the air filter or condenser and clean them if needed. Also, make sure the air ventilation screens aren't blocked. Ensure that the heat load isn't exceeding the chiller's capacity. High ambient air temperature or incorrect line voltage could also be factors; address these as necessary.
  • S
    Shawn GilmoreAug 5, 2025
    What to do if there is no fluid circulation in PolyScience Chiller?
    • R
      Robert ContrerasAug 5, 2025
      If your PolyScience Chiller isn't circulating fluid properly, start by checking the fluid level in the reservoir and add more fluid if it's low. Then, inspect the circulating system for any blockages and remove them. If the pump isn't running, it might need to be replaced.
  • M
    Michael KimAug 10, 2025
    Why does the pump cycle on and off at startup on my PolyScience Chiller and EFL flashes on display?
    • L
      Lauren BrownAug 11, 2025
      If the PolyScience Chiller pump cycles on and off at startup and 'EFL' flashes on the display, it indicates a no-flow condition. This is often due to insufficient fluid in the reservoir or a blockage in the circulating system. Add fluid to the reservoir as needed and remove any blockages.
  • W
    Wesley LucasAug 15, 2025
    How to improve insufficient circulation in PolyScience Chiller?
    • M
      Michele SparksAug 16, 2025
      If your PolyScience Chiller has insufficient circulation, the fluid viscosity might be too high; consider replacing it with a lower viscosity fluid. Also, check if the external tubing diameter is too small and replace it with larger diameter tubing if necessary. Restrictions in the fluid lines or low line voltage could also be contributing factors; check and correct these as required.
  • C
    Christopher JohnsonAug 20, 2025
    What to do if PolyScience Chiller does not run, but three decimal points appear on the temperature display?
    • D
      Daniel ParksAug 20, 2025
      If your PolyScience Chiller unit does not run and three decimal points appear on the temperature display (two on pressure/flow rate), it's likely in Standby mode. Press the Power Button on the front panel to exit Standby mode.
  • D
    David MillerAug 26, 2025
    How to fix a PolyScience Chiller that does not run and the digital displays are blank?
    • J
      Jeremiah WilliamsAug 26, 2025
      If the PolyScience Chiller isn't running and the digital displays are blank, ensure the electrical cord is securely connected to a working outlet. Also, check that the Power Switch/Circuit Breaker on the rear of the unit is turned ON.

Summary

Introduction and General Information

General Safety Information

Overview of safety symbols and essential precautions for operating the chiller.

Safety and Handling Procedures

Safety Recommendations, Unpacking, and Compliance

Essential safety advice, unpacking instructions, and regulatory compliance information.

Controls and Components Overview

Front Panel Interface

Detailed identification of controls and displays on the chiller's front panel.

Rear Panel Connections

Identification of power, circuit breaker, and optional serial ports on the rear.

Installation and Startup Procedures

Site Requirements and Electrical Setup

Guidelines for installation site, clearance, electrical power connection, and RS232 option.

Plumbing and Fluid Management

Process Piping and Fluid Filter Installation

Instructions for connecting process lines, selecting tubing, and installing a fluid filter.

Startup and Coolant Guidelines

Process Coolant Selection and Warnings

Guidance on choosing compatible fluids and warnings against using incompatible ones.

Normal Operation and Settings

Operational Parameter Configuration

Error and Fault Message Interpretation

Understanding Displayed Codes

Explains error/fault codes, their meanings, and required actions for resolution.

Routine Maintenance Procedures

Troubleshooting Common Issues

Problem Diagnosis and Corrective Actions

A guide to identifying problems and implementing solutions for chiller malfunctions.

Technical Specifications - 60 Hz Models

Technical Specifications - 50 Hz Model

RS232 Remote Communication

PolyScience LS Series Specifications