EasyManua.ls Logo

PR 4104 - Page 14

PR 4104
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 4104 V100-UK
Passive 2-wire mA:
Programmable ranges ................................. 0-20 and 4-20 mA
Direct or Inverted action
V-curve function, 100-0-100% ................... 20-0-20 mA
External 2-wire loop supply ....................... 3.5 - 26 V
Common specifications current output:
Signal range ................................................ 0...23 mA (unipolar) /
-23...+23 mA (bipolar)
Current limit ................................................. ≤ 28 mA (unipolar) /
±28 mA (bipolar)
Load stability ............................................... ≤ 0.005% of span / 100 Ω
Programmable damping .............................. 0.0...60.0 s
Output limitation at outside range:
on 4...20 and 20...4 mA signals .............. 3.8...20.5 mA
on other unipolar signals ......................... 0 and 115% of max. value
on bipolar signals .................................... ±115% of min. and max. values
Sensor error indication, at 4...20 mA input:
selectable ................................................ Low, High, Zero or None
Low - corresponds to 0 mA at 0...20 mA and to 3.5 mA at 4...20 mA
High - corresponds to 23 mA at both 0...20 and 4...20 mA
Zero - equals 0 mA output
None - the output state is undefined
Voltage output specifications:
Programmable ranges ................................. 0/0.2-1, 0/1-5, 0/2-10, ±1, ±5 and
±10 V - Direct or Inverted action
V-curve function, 100-0-100% ................... 1-0-1 V, 5-0-5 V and 10-0-10 V
Load, min .................................................... >500
Programmable damping .............................. 0.0...60.0 s
Output limitation at outside range:
on unipolar signals starting from 0 ......... 0 and 115% of max. value
on unipolar signals with offset ................ -5% of min. value and
115% of max. value
on bipolar signals .................................... ±115% of min. and max. values
Sensor error indication, at 4...20 mA input:
selectable ................................................ Low, High, Zero or None
Low - corresponds to the selected min range value
High - corresponds to the selected max range value
Zero - equals 0 V output
None - the output state is undefined