Trouble Shooting Guide
SYMPTOM
POSSIBLE ISSUES OR COMMENTS
SOLUTIONS
Module will not fit into
slot
The Module is designed to be inserted one
way only
Rotate the Mod
ule 180 degrees and it
will fit
Unit Does Not
Turn On
Module is missing or not fully inserted
Batteries are not property installed
Batteries are dead or are missing
Insert or Re-insert module properly
Re-orient batteries properly
Replace batteries
Difficult to Peel Off the
Pads Liner
Peel the liner from the Tab End
Gently bend over the Connector to start
separation
Peel with thumb if necessary
Remove and throw away liner
Refer to the Assembly Section
Both pads attached to
Manikin but
device will not advance to
Analysis
Operational solution
Pad is inserted upside down in the cable
connector.
Adhesive worn out
Insure Pads are being pressed on
firmly
Flip pad around and secure with black
pad pin
Replace pad
Pad pulls out from pad
connector
Check for black pad pin missing Install new black pad pin
Lid will not stay closed
Pads Case is not placed in proper position
Re-seat pads case
Wires not properly stored in compartment
under Case - Wire may be caught under
edge of Pads Case
Re-coil the wires
Pads Case Front & Back not properly closed
Snap case together on all 4 corners
Device Proceeds to
Analyzing before
both pads have been
placed on the
manikin
Pads not properly positioned in Case
Look for pad riding up pad case wall
Re-position pad in case to lay
completely flat so it is not touching
side walls