EasyManua.ls Logo

Proactive Protekt Aire 2000 - Maintenance Trouble Shoot; Technical Specifications; Package Contents and Specifications

Proactive Protekt Aire 2000
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAINTENANCE TROUBLE SHOOT / TECHNICAL SPECIFICATIONS
Trouble Signal Failure Cause Fault Handling
No air output
Power indicator
is ON.
The inflation tube is not con-
nected to the control unit.
Connect the air tube.
The silicone tubes inside the
control unit are disconnected.
Contact Service Personnel.
Power indicator
is OFF.
No power. Turn on power switch
I
ncomplete inflation
overlay too soft.
Small hole at the air cell, or
overlay life comes to its end.
Replace the air cell or
the whole overlay.
Air knob indicate at “MIN”
location.
Turn the air tube at the
right to give max pressure .
PACKAGE CONTENT LIST
Control Unit x 1 Overlay Unit x 1 • User Manual x 1
TECHNICAL SPECIFICATIONS
INPUT RATING: 120VAC, 60Hz, 1A FUSE RATING: 250V, 1A
Control Unit Overlay
Item #: 80021 Item #: 80022
Power Supply: 120V/60Hz Size: 80’’L x 36’’W x 5’’H
Air output: 8 liter/min Top Cover Material: Nylon/PU
Pressure range: 30-60 mmHg Air Cells Material: Nylon/PVC
Cycle Time: 10 min Bottom cover material: Nylon/PVC
Size: 11’’ x 5.9’’ x 3.5’’ # of aircells: 20
Weight: 1.4kg (3lbs) CPR Valve? YES
9

Related product manuals