Promax DC-800 - Setting the Parameters for the Optional Settings

Promax DC-800
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Press the Right Arrow key to move to the next setting. The Soft Air LED will remain on.
However you are now setting the Vent Off time.
4. Using the Plus/Minus keys set the amount of time you want the vent valve to remain
closed between each vent pulse.
5. Press the Up Arrow key to return to normal operation mode.
3.4.2
Setting the Parameters for the Optional Settings
ImportantNote: Thesesettingswillonlyfunctioniftheoptionstheypertaintowerepurchased.
Refertofigure6belowtohelpyoucompletethesesettings.
Figure
6
PRO G RA M
PROGRAM TI ME /
%
SK IP
CYCLE
STO P
CYCLE
STB
W
ORK
VAC
GAS
SEAL1 SEAL2
COOL
SO FT
GAS
SH IFT
GAS
ON/OFF
PU LSE
ON/OFF
O IL
CHG
AIR
ON/OFF
PU LSE
PU R G E
PO W
ER
ON
PO W
ER
OFF
Gas On / Off /Pulse LED
Purge Gas On / Off Pulse On / Off
A. Gas Flush
1. Prepare to make changes as instructed in point D
2. Press the Left Arrow key until the Gas LED is lit.
3. Using the Plus /. Minus keys set the amount of time you want gas to enter the
chamber.