Overvoltage Category I (CAT I):
Equipment for connection to circuits in which measures are taken to
limit the transient overvoltages to an appropriate low level.
Overvoltage Category II (CAT II):
Energy-consuming equipment to be supplied from the fixed
installation.
Overvoltage Category III (CAT III):
Equipment in fixed installations, distribution boards, and circuit breakers.
Overvoltage Category IV (CAT IV):
Origin of installation or utility level measurements on primary over-current
protection devices and on ripple control units.
SAFETY INFORMATION: (Read First Before Operation)
Please follow the following instructions carefully for safe operation.
NEVER use the clamp meter for Voltages higher than 600V.
DO NOT hold the clamp meter beyond its tactile barrier.
DO NOT use the clamp meter and accessories if they look
damaged.
USE CAUTION when working with high voltages.
USE CAUTION when measuring the voltages higher than
30VAC rms or 60VDC. These voltages pose a shock hazard.
USE EXTREME CAUTION when working around bare
conductors or bus bars.
ALWAYS use the clamp meter as the instructions in the manual.
WARNING: If the clamp meter is used in a manner
Not specified by the manufacturer, the protection
Provided by the clamp meter may be impaired.