Performance verification 5
Conduct the performance verification to check the performance of the Puritan Bennett™
840™ Ventilator System. Field performance verification testing includes electrical safety tests,
the extended self test (EST), plus specific performance verification tests, using the
Puritan Bennett PTS 2000 Performance Test System with Puritan Bennett 840 VTS software. A
manual ventilator check can be performed to check general ventilator performance.
NOTE:
Performance verification procedures do not apply to ventilator accessories. Refer
to the specific accessory operator’s or service manual for performance verification
information. Malfunctioning accessories may affect some ventilator functions and
may result in false test results.
5.1 Tools, test equipment, and service materials
In addition to the standard set of tools listed in Chapter 1 of this manual, the test
equipment and materials listed in Table 5-1 are required for performance verification
procedures. Specific parts required for conducting the performance verification,
using the PTS 2000 Performance Test System with BreathLab
®
840 Ventilator Test
Software (VTS), are indicated with “X” in the third column.
5.2 When to run
Run the entire performance verification after servicing or repairing the ventilator. The left-
hand column of Table 5-2 lists the types of service activity that require performance
verification. In addition, run the electrical safety tests, which are part of the performance
verification, in accordance with your hospital’s requirements.
Table 5-1: Tools, equipment, and service materials required for
performance verification
Description
Manufacturer/model or
Puritan Bennett
part number
Required for
PTS 2000
use?
Puritan Bennett 840 VTS (Ventilator Test Software) 4-075359-00 and
current revision Software
Download CD-ROM
X
Adapter, patient pressure (“T” connector) *4-011521-00 X
Cable, trigger *4-075360-00 X
Cable, null modem *4-075361-00 X
* These parts may be purchased individually or by ordering a Puritan Bennett 840 VTS Accessories Kit, P/N 4-076599-00.
Puritan Bennett 800 Series Ventilator System Service Manual