PROBLEM CAUSE REMEDY
No pressure in the cuff. An adequate pressure
cannot be generated in
the cuff. A leak may have
occurred.
1. Check that the cuff is correctly positioned and fits on the arm.
2. Replace the batteries, if necessary. Repeat the measurement.
If the problem occurs again, contact customer service.
Abnormal result. The measuring signals
are inaccurate therefore
no result can be
displayed. This could be
due to the cuff not being
fully deflated before
measurement, noise
interference, user talking,
user movements, cuff
not correctly fastened,
cuff broken, pump or
valve failure, pressure
overflow, or other special
characteristics of the user.
1. Read through the instructions for performing reliable
measurements. Re position the cuff and repeat the
measurement while keeping your arm still and without
talking. If the problem occurs again, please contact
customer service.
2. If user has special conditions, please contact your physician.