Manufacturer’s declaration-electromagnetic immunity
The QardioArm (Model No.: A100/A101) Wireless Blood Pressure Monitor system is intended for use in the electromagnetic environment
(for home healthcare) specified below. The customer or the user of the QardioArm Wireless Blood Pressure Monitor system should assure
that is used in such and environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-guidance
(for home healthcare environment)
Conducted RF
IEC 61000-4-6
3 Vrms: 0,15 MHz – 80 MHz
6 Vrms: in ISM and amateur
radio bands between 0,15
MHz and 80 MHz
80 % AM at 1 kHz
Not applicable
Not applicable
Portable and mobile RF communications
equipment should be used no closer to any part of the QardioArm
Wireless Blood Pressure Monitor system including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Recommended separation distance:
d = 1,2 √P √P
d = 1,2 √P √P 80MHz to 800 MHz
d = 2,3 √P √P 800MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.