7
INSTALLATION
WARNING!
x High voltage is present in the generator, convertor and high frequency cable. Do not
attempt
to remove the generator cover or convertor case.
x Do Not
touch any open cable connections on the unit while the power is turned on.
A. Power Requirements
The generator requires a single phase, grounded, three-wire, 115V or 220V, 50/60 Hz source,
unless otherwise fitted, and has a 3 amp fuse for 115V service or a 1.5 amp fuse for 220V
service. There is a select switch on the rear panel of the generator to set voltage for 115V (92 ±
140RMS) or 220V (200 ± 240 VRMS) service.
CAUTION: Do not
operate a unit set for 115V service on 220V line or operate a unit set for
220V service on a 115V line.
WARNING: The electrical line cord is equipped with a 3-prong grounding plug. Do not
remove
the grounding prong under any circumstances. The plug must be plugged into a mating 3-prong
grounded outlet.
NOTE: Only use IEC approved fuses. Do not
XVH³VORZEORZ´IXVHVRUIXVHVUDWHGDERYHWKH
amperage noted.
Cleaning Instructions
The generator and convertor may be cleaned using Windex or a similar acid-free cleaning
solution, and an anti-static cleaning cloth. Probes can be cleaned with isopropyl alcohol and
sterilized in an autoclave (probes only).